This Waters 2414 RID has been Refurbished, Fully Tested, Inspected and Verified to be in Great Working Condition.
Warranty: 90 Days
Includes: Waters 2414 RID & Power Cord
Model: 214
Serial: A15214943M
About:
Designed to provide scientists with the highest sensitivity, stability, and reproducibility, the 2414 Refractive Index (RI) Detector is the ideal solution for the analysis and quantification of components with limited or no UV absorption, such as alcohols, sugars, saccharides, fatty acids, and polymers.
With the 2414 RI Detector, you can detect components without UV chromophores, achieve highly reproducible detection with a wide linear dynamic range, and attain stable thermal control with temperature equilibration of incoming solvent and a thermally isolated optics bench. The 2414 RI Detector completes a total sample-centric solution for many applications, including polymer analysis, quality control (QC) applications, preparative analysis, analytical analysis, and food applications.
Specifications:
|
Refractive index range |
1.00 to 1.75 RIU |
|
Measurement range |
5.0 x 10-4 RIU to 7.0 x 10-9 RIU |
|
Linear dynamic range |
≤5.0% over ±5.0 x 10-4 RIU |
|
Noise2 |
±1.5 x 10-9 RIU (2 s FTC Hamming 1.0 mL/min, 100% H2O) ±3.0 x 10-9 RIU 410/2410 emulation mode (1 s FTC (RC) 1.0 mL/min, 100% H2O) |
|
Drift |
2.0 x 10-7 RIU/hour |
|
Filter time constant |
0.0 to 5.0 s (Hamming) 0.0 to 10.0 s (RC) |
|
Compatible flow rate range |
0.1 to 10.0 mL/min |
|
Attenuation settings |
1 to 500 x 10-6 RIU 1 to 1024 maximum in emulation mode |
|
Temperature control |
Internal oven: 30 to 55 °C (86 to 131 °F), ±0.5 °C, settable 1 °C increments One external column heater, steel: ambient to 150 °C (302 °F), ±1.0 °C, settable 1 °C increments |
Product ID # 2709