This Beckman Coulter AU480 has been fully cleaned, decontaminated and inspected.
30 Day Warranty!
Includes:
Beckman Coulter AU480
- Computer with Beckman Dongle, Monitor, Keyboard, Mouse,
- Drain Hoses
- 3x Washer Solution Containers, Cleaner and Bottles
- Power Cord
- Everything Shown
- ELGA Medica pro water purifier system also available. Please inquire
Year: 2017
Count: 814,662
About from Beckman:
Robust testing capabilities
- Broaden your test offerings with more than 120 different Beckman Coulter system reagent applications available
- Conduct quick priority testing using the AU 480 clinical chemistry analyzer’s STAT wheel
- Turn around critical samples fast with sample pre-dilution and auto-repeat of over-range results
Efficient reagent and consumable use
- Minimize hands-on time and reduce daily reagent inventory tasks with ready-to-use liquid concentrated reagents
- Experience lower total cost of ownership with low consumable requirements
Convenient operational features
- Provide the right operator support when needed using context-sensitive help screens and instructional videos
- Simplify operations with user-friendly graphical user interface design and touch-screen monitor
- Reduce tech-to-tech variation and achieve consistent result quality with AU family of chemistry analyzers' standardized processes
Analytical System | Fully automated, random access chemistry system with STAT capability |
Analytical Principle | Spectrophotometry and potentiometry |
Analytical Types | Endpoint, rate, fixed point and indirect ISE |
Analytical Methods | Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE |
Simultaneously Processed Analytes | 60 photometric tests + 3 ISE, 120 pre-programmed onboard tests |
Throughput | 400 photometric tests/hour; maximum of 800 with ISE |
ISE Throughput | 200 samples/hr. Maximum tests/hr: 600 if ISE only |
Sample Types | Serum, plasma, urine, other |
Sample Feeder | Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 80 samples; continuous loading |
Sample Tubes | Primary and secondary tubes; diameter from 11.5 to 16 mm; height from 55 to 102 mm, nestled micro sample cups |
STAT samples | Up to 22 positions for STAT samples via cooled turntable. Easy-to-operate interrupts between tests. Auto-repeat run capability. |
Sample Volume | 1–25 μl in 0.1 μl steps (1-25 μl for repeats) |
Reagent Capacity | 76 positions for (R1+R2, detergent position). Holds 15, 30 and 60ml bottles. |
Reagent Storage | Refrigerated (4–12°C) |
Reagent Volume | R1: 10–250 μl; R2: 10–250 μl (in 1 μl increment) |
Total Reaction Volume | 90–350 μl |
Reaction Cuvette | Permanent glass cuvettes |
Reaction Time | Up to 8 minutes, 37.5 seconds |
Reaction Incubation | 37°C, dry bath |
Photometry System | Direct assay through the reaction cuvette (0–3.0 OD) mono and bichromatic measurements possible |
Wavelength | 13 different wavelengths between 340–800 nm |
Calibration | Auto calibration, cooled calibrator positions; master calibration established by two-dimensional barcode |
Quality Control | Auto QC, cooled QC positions |
Rerun | 1. Auto rerun and manual rerun available |
Sample Pre-dilution | Dilution ratio: Fixed to 3, 5, 10, 15, 20, 25, 50, 76 or 100 |
Test Requisition | Individual and profile test requisition via online, mouse, function keys and touch-screen |
Sample Integrity | Lipaemia, Haemolysis and Icterus analysis. Sample clot detection and probe crash protection. |
Online | RS232C. Full uni- and bidirectional communication possible. |
Software | Windows 7® |
Data Storage | Sample number 100,000 samples. Reaction Data 200,000 tests. |
Dimensions (W x H x D) | ANL: 1450 mm x 1205 mm x 770 mm |
Power Supply | 100–240v; 60 Hz/ < 3.5 kVA |
Product ID # 2676