This Thermo Dionex UltiMate 3000 HPG-3200BX has been fully tested, inspected and verified to be in great working condition.
Warranty: 90 Days
Type: Biocompatible Binary Semi-Preparative Pump
Model: HPG-3200BX
Part No: 5042.0025
Includes: Thermo Dionex HPG-3200BX Binary Pump, Power Cord & Inlet Hoses
Recommended Use: Purify biological compounds in the microgram to lower-milligram range quickly and accurately with the Thermo Scientific™ UltiMate™ HPG-3200BX Biocompatible Binary Semi-Preparative Pump. Binary high-pressure proportioning offers gradient accuracy and precision. Maximize your flow rate depending on your application. This pump supports operating pressures of up to 16.5 MPa (2400 psi/165 bar) at up to 30mL/min.; 13.5 MPa (1950 psi/135 bar) at 50mL/min. For isocratic applications, Double Flow mode enables flow rates of up to 100mL/min.
Features:
- Serial dual-piston operating principle for outstanding flow rate accuracy and precision across the entire flow rate range
- Wide operating flow rate range from 0.5 mL/min to 50 mL/min
- Support operating pressures of up to 165 bar (2400 psi) at up to 30 mL/min; 135 bar (1950 psi) at 50 mL/min
- Double Flow mode enables flow rates of up to 100 mL/min for isocratic applications
- Binary high-pressure proportioning for best-in-class gradient accuracy and precision
- SmartFlow™ technology automatically ensures optimal chromatographic performance at any operation condition
- Automatic compressibility compensation, no manual solvent setting needed
- Active rear-seal wash and floating pistons for maximum seal lifetime
Specifications:
Flow Accuracy | ±0.1% at 3 mL/min typically ±0.1% (at 25 mL/min) |
Flow Precision | <0.05% RSD or <0.01 min SD, whichever is greater |
Gradient Formation | High-pressure proportioning |
Max. Flow Rate | 50.0 mL/min. |
Pressure Range | 2 to 16.5 MPa up to 30 mL/min; 2 to 13.5 MPa up to 50 mL/min |
GLP Features | Full support of Thermo Scientific Dionex AutoQ instrument qualification, qualification status and system wellness monitoring. All system parameters are logged in the Chromeleon Audit Trail. |
Gradient Delay Volume | 800-1550 µL Default: 800 µL |
I/O Interfaces | 2 digital inputs, 2 relay outputs |
Min. Flow Rate | 0.001 mL/min. |
No. of Eluent Lines | 2 |
Operating Principle | Serial dual-piston |
Pressure Ripple | Typically: <0.2 MPa or <1% whichever is greater |
Proportioning Accuracy | ±0.2% (full scale) |
Proportioning Precision | < 0.2% SD |
Unit Size | Each |