Steris Amsco Century Steam Sterilizer V-136H

Steris Amsco Century Steam Sterilizer V-136H

$7,995.00

This sterilizer is in good working condition. 60 Day Warranty at asking price!

Model: V-136H / SV-136H

Interior Dimensions: 26W x 37.5H x 45 inches (racks are 36" deep)

Exterior Dimensions: 60W x 60D x 72-76H inches

Steam Source: Building Steam

Electrical

Vacuum Pump: 208/240 Volt, 3-Phase, 60 Hz, 6 Amps per phase, or 480 Volt, 3-Phase, 60 Hz, 3 Amps per phase.

Control: 120 Volt, 1-Phase, 60 Hz, 2 Amps

Hinged Door :  Left Hinge : Opens on Right

The sterilizer is a prevacuum configuration and designed for fast, efficient sterilization of heat- and moisture- stable materials in scientific applications. The prevacuum configuration sterilizer is equipped with prevacuum, gravity, liquid, leak test, and daily air removal (Bowie-Dick) test cycles.

Features

- 6 x 37.5" (660 x 950 mm) chamber cross-section is sized to allow for efficient, high-volume processing.

- Resistance Temperature Detectors (RTDs) are installed for sterilizer temperature control. The chamber drain line RTD senses and controls temperature variations within the sterilizer chamber. A jacket RTD provides temperature control within the jacket space. These RTD signals, converted into electrical impulses, provide accurate control inputs and readouts throughout the entire cycle.

- Automatic utilities startup/shutdown permits slow cooling of entire vessel and load. Shutdown may be programmed to activate at end of any designated cycle or time of day. When activated, control system automatically shuts off all utility valves, conserving steam and water usage. Sterilizer utilities can be restarted either by programmed time or manual operation. A different shutdown and restart time can be programmed for each day.

-Insulation, one-inch thick, asbestos-free spin-glass (rated at 500 °F [260 °C] continuous) encompasses the exterior of the sterilizer vessel and is sealed in an oil and water resistant outer jacket.

Vacuum System 

Vacuum pump reduces chamber pressure during prevacuum and post-drying phases. Air is drawn from chamber through vacuum system. Following dry phase, chamber vacuum is relieved to atmospheric pressure by admitting air through a bacteria-retentive filter.

Steam Source

Sterilizer is piped, valved, and trapped to receive building- supplied steam delivered at 50 to 80 psig (3.5 to 5.6 bar) dynamic. Steam piping is constructed of brass and includes a shutoff valve, steam strainer, and a brass pressure regulator.

Steam feeds from jacket to chamber. A check valve is added between jacket and chamber on sterilizers with decontamination cycle option.

Steam Consumption

Avg. 112 lb/hr (51 kg/hr) 

Peak: 190 lb/hr (86 kg/hr

FOR MORE INFORMATION, CLICK HERE FOR DATA SHEET